What this test observes
Ultraviolet examination records a specimen’s visible response to a stated excitation source. Fluorescence is seen during excitation; phosphorescence persists after that source is removed. Responses can vary within a species.
Underlying property: Light and colour.
Parts and setup
- Identify the lamp, wavelength setting and protective viewing enclosure. This module compares LWUV at 365 nm and SWUV at 254 nm.
- Use the instrument’s protective viewing arrangement and operating guidance; shield eyes and skin from direct exposure, especially with SWUV.
- Observe in a suitably dark environment. Record source wavelength, response colour, strength and distribution.
- Remove the excitation and observe whether a visible afterglow remains. Compare with other identification evidence.
Explore the instrument in 3D
Drag to orbit the instrument. Select a part on the model or use the part buttons to learn what it does.
Preparing the 3D instrument…
Choose a part to inspect.
Original schematic teaching model. External proportions and internal placement are schematic, not a manufacturer’s dimensional drawing. The observation lab below shows the separate viewing result.
Make an observation
Interpretation and limitations
Compare fluorescence under the stated LWUV and SWUV conditions and record afterglow separately. An inert response does not identify a material. Fluorescence colour alone cannot establish natural or synthetic origin, geographic origin or a treatment. Lamp output, viewing filters, contamination and specimen chemistry can alter observations.
Check your understanding
Answer each question before opening its explanation.
1. How do fluorescence and phosphorescence differ?
Fluorescence is observed during excitation; phosphorescence continues after the excitation is removed.
2. Why specify LWUV or SWUV?
Different excitation wavelengths can give different responses in the same specimen.
3. Can a fluorescence colour establish natural origin?
No. Responses overlap and vary; independent evidence is needed.