What this test observes
A spectroscope spreads transmitted light into a spectrum so that selective absorption can be observed as dark bands or lines. The light source, alignment, slit and specimen all affect what can be resolved.
Underlying property: Dispersion and selective absorption.
Parts and setup
- Identify the viewing lens, slit and dispersing element: prism or diffraction grating.
- Arrange a suitable continuous white source and align light passing through the specimen with the slit. Keep stray light out of the view.
- Focus the slit where the instrument permits, then balance brightness against resolution.
- Compare features with a reliable reference, checking the spectrum’s orientation and any calibrated wavelength scale.
Explore both 3D cross-sections
Both instruments separate wavelengths. The grating uses diffraction and interference; the prism train uses dispersion. Their spectrum spacings differ.
Diffraction grating spectroscope
Light enters at the left; the eye views from the right. Drag to orbit, or select a numbered component to inspect its role.
Preparing the 3D cross-section…
Choose a part to inspect.
Prism spectroscope
Light enters at the left; the eye views from the right. Drag to orbit, or select a numbered component to inspect its role.
Preparing the 3D cross-section…
Choose a part to inspect.
Original schematic models informed by The Gemology Project’s spectroscope diagrams. Component shapes, prism angles, grating rulings and ray paths are illustrative, not a manufacturer’s construction drawing or a calculated optical trace.
Make an observation
Absorption and emission spectra
Absorption removes wavelengths from a continuous source, leaving dark bands or lines. Fluorescence adds bright emission lines under suitable excitation. It is not simply an inverted absorption pattern. Ruby and red spinel can show bright red features; a sodium source gives a bright yellow line.
Compare the source alone with light from the specimen. A line-rich lamp can introduce its own features. The red doublet of ruby and yellow sodium doublet are usually unresolved in a handheld view; the bright/dark appearance also depends on illumination.
Reference features
All wavelength values below are approximate and given in nanometres (nm). Confirm an interpretation with other gemmological properties. Cobalt absorption alone does not distinguish natural from synthetic spinel.
| Example | Contributor | Absorption | Emission |
|---|---|---|---|
| Ruby | Chromium | Red doublet near 694; broad yellow–green absorption around 540–580; blue lines near 468 and 476; violet absorption. | Red fluorescence near 694; the close doublet usually appears unresolved. |
| Red spinel | Chromium | Principal red lines near 675 and 686; broad green absorption; ruby-like fine blue lines are absent. | An organ-pipe group of red lines may fluoresce. Only the two principal reference positions are plotted. |
| Emerald | Chromium (iron may contribute) | Red doublet near 680/683, further lines near 637/646/662; broad 580–620 band; blue lines near 472.5/477.5; violet absorption. | No emission pattern plotted in this example. |
| Almandine garnet | Iron | Three broad absorption bands centred near 505, 525 and 575; bands can merge. | No emission pattern plotted in this example. |
| Spessartine garnet | Manganese; iron may contribute | Violet lines around 412, 424 and 432; further features near 460 and 480. Iron bands may also occur. | No emission pattern plotted in this example. |
| Iron-bearing blue sapphire | Iron-related absorption | Strong blue feature near 450, with weaker features sometimes described near 460 and 471. | No emission pattern plotted in this example. |
| Iron-bearing blue spinel | Iron | Blue lines near 458 and 478, with weaker bands around 555 and 585. | No emission pattern plotted in this example. |
| Cobalt-coloured synthetic blue spinel | Cobalt | Three broad bands around 540, 580 and 635 in this representative pattern. | No emission pattern plotted in this example. |
| Zircon | Characteristic zircon pattern | A strong fine line near 653.5; additional lines may occur. | No emission pattern plotted in this example. |
| Cape-series yellow diamond | Nitrogen-related defects | A fine violet feature near 415, with other faint features possible. | No emission pattern plotted in this example. |
| Peridot | Iron | Three bands in the blue to blue–green; bands may be faint, edge-like or merged. | No emission pattern plotted in this example. |
| Sodium light source | Sodium — source example | A light-source example, rather than a gemstone absorption pattern. | Bright yellow doublet near 589.0/589.6; usually one unresolved line in a handheld instrument. |
Interpretation and limitations
A narrow slit can improve separation but reduces the light admitted. A broad slit can merge nearby features. Prism and grating layouts distribute colours differently; do not compare positions as if their scales were identical. Absence of a visible feature may reflect technique, weak absorption or inadequate illumination. Not every stone has a useful visible spectrum.
Check your understanding
Answer each question before opening its explanation.
1. What is the slit-width trade-off?
A narrower slit can improve resolution while reducing admitted light.
2. Are prism and grating colour spacings identical?
No. The red end is relatively compressed in the prism view.
3. Does this practice spectrum give a species identification?
No. It teaches observation using unnamed, uncalibrated patterns.
4. How do absorption and emission lines differ?
Absorption features are dark where light is removed. Emission features are bright where excited material adds light; not every absorption pattern has a corresponding visible fluorescent pattern.
5. Does a cobalt spectrum prove that blue spinel is synthetic?
No. Natural spinel can also be cobalt-coloured. Spectra require interpretation with other evidence.